Abstract

A novel platinum nanoparticle-modified graphite pencil electrode (PtNP-GPE) for a non-enzymatic determination of H2O2 is proposed. The PtNP-GPE is prepared by heating a GPE in an aqueous solution of ammonium tetrachloroplatinate(II) and ascorbic acid at 75°C for 15min. Both unmodified and PtNP-GPE are characterized using field-emission-SEM and cyclic voltammetry. The PtNP-GPE showed better electrocatalytic properties toward the electrochemical redox reaction of H2O2 than that of the unmodified GPE. The PtNP-GPE amperometric response is tested at four different potentials (+0.7, +0.5, −0.2 and −0.3V vs. Ag/AgCl reference electrode). Taking into consideration both the intensity of the amperometric response and detector stability, the applied potential of +0.5V is the best among the tested potentials. The PtNP-GPE amperometric response of H2O2 at +0.5V is linear from 10 to 110μM (R2=0.999) with a detection limit of 3.6μM.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.