A glassy carbon electrode modified with electrochemically reduced graphene oxide (ErGO) and poly(3,4-ethylenedioxythiophene):poly(styrenesulfonate) (PEDOT:PSS) was developed for uric acid determination with dopamine as interference in artificial saliva. The electrochemical performance of the fabricated electrode was studied using both techniques of cyclic voltammetry (CV) and differential pulse voltammetry (DPV), under optimized conditions. Using DPV, the sensor based on ErGO/PEDOT:PSS modified GCE displayed a linear relationship in the concentration ranges of 10-100µM for uric acid. This uric acid sensor exhibited a high sensitivity with detection limits of 1.08µM and quantitation limit of 3.61µM. This sensor also showed good reproducibility for uric acid detection in artificial saliva in 5 consecutive days of measurements. This device was successfully used to analyze uric acid in artificial saliva as well as in a human saliva sample using the standard addition method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.