An rf-driven reaction cell offers the potential for enormous efficiency in the removal of interfering ions prior to mass analysis in inductively coupled plasma mass spectrometry. Concomitant with this efficiency is the potential for interference production within the reaction cell. Operation of a low rf-amplitude quadrupole reaction cell in a bandpass mode, either rf-only or with additional dc voltage between pole pairs, is shown to suppress these in-cell interferences. Examples of various types of interferences (spectral and continuum) are shown, and their elimination without sacrificing sensitivity is demonstrated. The advantage of dynamically sweeping the bandpass in concert with the mass analyzer is shown. The necessity of efficiently evacuating the reaction cell (in this case by venting it to the high vacuum chamber) when reaction gas is not added is demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.