Abstract

Microemboli passing to the cerebral circulation during cardiopulmonary bypass can contribute to postoperative neurologic dysfunction. Many studies conclude that air microbubbles predominantly are responsible for this problem. A dynamic bubble trap (DBT) was developed to diminish the number of microbubbles in the arterial line of extracorporeal circulation. The DBT is able to substantially reduce the number of air microbubbles, as shown in two patients undergoing coronary artery bypass grafting, where a high number of microbubbles was assessed. Although a 40-µm arterial filter was used, many bubbles larger than 40 µm occurred in the arterial line. The DBT reduced the number of large microbubbles from 2,267 to 67 in patient 1 and from 897 to 61 in patient 2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.