Abstract

A new current-comparator-based dual-channel calibration system for AC currents and AC voltages in the millivolt range has been developed. The range of generated currents is 1 mA–5 A, and the range of generated voltages is 1–500 mV. The calibration system in-phase and quadrature expanded measurement uncertainties ( $k=2$ ) are within $25\cdot 10^{\mathrm {\mathbf {-6}}}$ . Although the system has been developed primarily for operation at power frequencies, it is also capable of operating at frequencies up to 1000 Hz with slightly increased uncertainties. The traceability to SI units is established through the calibration of AC voltage by means of AC/DC transfer and DC voltage reference, standard resistor, capacitor, and current comparator ratios.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.