To compare setup uncertainties at two different institutions by using identical imaging and analysis techniques for thoracic patients with different fixation equipments. Patient registration results from Cone-Beam CT (CBCT) scans of 174 patients were evaluated (1068 CBCT scans). Patients were fixated using a standard or custom made fixation at Royal Marsden Hospital and Odense University Hospital, respectively. Five imaging protocols were retrospectively simulated to compare the fixation equipments. Systematic and random setup uncertainties were calculated to estimate sufficient treatment margins. The setup uncertainties are of similar sizes at the two institutions and there is no observable drift in the precision of the fixation equipments during the treatment course. When a correcting imaging protocol is performed there is a significant increase of the systematic setup uncertainties in between imaging fractions. A margin reduction of ≥0.2 cm can be achieved for patients with peak-to-peak respiration amplitudes of ≥1.2 cm when changing from 4D-CT to Active Breathing Coordinator™ (ABC). The setup uncertainties at the two institutions are the same despite different fixation equipments. Hence margins cannot be reduced by changing fixating equipment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.