Abstract

Based on the parametric oscillation process, we demonstrate the dual-axis phase-sensitive demodulation (PSD) and single-axis phase-insensitive demodulation (PISD) for the atomic magnetometer in an in situ nuclear magnetic resonance (NMR) detection system, which can separate the precession signals of NMR from the oscillating magnetic fields. The two orthogonal magnetic fields can be detected simultaneously and independently by selecting the optimal demodulation phases with the traditional PSD method. The response signals of the parametric modulation magnetometer demodulated with high order harmonic signals are evaluated, which is a new exploration. The first order harmonic demodulation can present the best sensitivity about 250 fT/Hz1/2. The high order harmonic demodulation technology supplies a twofold 3 dB bandwidth. With the PISD method, a single-axis demodulation technique is proposed. The transverse nuclear spin precession magnetic fields can be extracted effectively with the demodulation R signal outputs by setting a specific longitudinal modulation magnetic field amplitude, which is a new demodulation strategy compared with the traditional demodulation method for the NMR system.