Abstract
In most cases calibration of measuring instruments is obtained by comparing measurement results to a measurement standard. But in the case of coordinate measuring system, the availability of reliable measurement standards is often limited: calibrated ball plated, for example, are expensive and difficult to maintain. This originated the so called “self-calibration” procedures, in which an uncalibrated artifact is adopted to estimate the stage error of the system.One of the most widespread coordinate measuring systems is an XY stage. Several procedures have been proposed for the self-calibration of this kind of system. Most of them are based on performing three or more measurements of an uncalibrated plate, with the position of the plate changing from one measurement to the other. Oftentimes, the positions are only slightly different from each other, thus making it difficult to obtain them with a good accuracy when measurements are carried out at a micro scale. In this work, an approach is proposed which allows larger displacement from one measurement to the others, thereby allowing an easier management for micro XYstages.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.