A simple double current source, designed for supplying two windings of a cryogenic current comparator (CCC) is described. The ratio of the two currents is preadjusted by means of an auxiliary resistor ratio and precisely regulated by a CCC associated with a SQUID detector. This enables accurate ratio calibration of four-terminal standard resistors at room temperature, taking advantage of high accuracy (better than 1 × 10-9) and resolution (5 × 10-9 A in a one-turn winding) of the CCC. Operational experience in the calibration of standard resistors (1:10 ratio) is reported. In addition, the use of this double source is suggested for the realization of a cryogenic voltage standard.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.