An experimental arrangement of a radon chamber with an intrinsic constancy of the relative radon concentration is described. The system consists of a reference chamber and an auxiliary storage chamber. The only active device is a timer-controlled pump or valve which feeds radon gas from the storage into the reference chamber. The switching pattern of the timer is extracted from model calculations and theoretically performs an exact compensation of the radon loss by radioactive decay. If the calculations are done in real time and online, every known external event influencing the radon concentration can be compensated. This paper presents a simple timer circuit and a computer code which generates the timer program. The influence of the air flow stability and the leakage of the chambers are discussed. It is planned to apply this theoretical approach to provide a constant radon gas concentration for an actual chamber.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.