Abstract

It is essential to construct a biofuel cell-based sensor and develop an effective strategy to detect glucose without any potentiostat circuitry in order to create a simple and miniaturized device. In this report, an enzymatic biofuel cell (EBFC) is fabricated by the facile design of an anode and cathode on a screen-printed carbon electrode (SPCE). To construct the anode, thionine and flavin adenine dinucleotide-dependent glucose dehydrogenase (FAD-GDH) are covalently immobilized via a crosslinker to make a cross-linked redox network. As a cathode, the Pt-free oxygen reduction carbon catalyst is employed alternative to the commonly used bilirubin oxidase. We proposed the importance of EBFC-based sensors through the connection of anode and cathode; they can identify a short-circuit current by means of applied zero external voltage, thereby capable of glucose detection without under the operation of the potentiostat. The result shows that the EBFC-based sensor could be able to detect based on a short-circuit current with a wide range of glucose concentrations from 0.28 to 30 mM. Further, an EBFC is employed as a one-compartment model energy harvester with a maximum power density of (36 ± 3) μW cm− 2 in sample volume 5 μL. In addition, the constructed EBFC-based sensor demonstrates that the physiological range of ascorbic acid and uric acid shows no significant effect on the short-circuit current generation. Moreover, this EBFC can be used as a sensor in artificial plasma without losing its performance and thereby used as a disposable test strip in real blood sample analysis.