Abstract

A disposable electrochemical immunofiltration test strip for the rapid detection of α-fetoprotein (AFP) was developed. The test strip was constructed by assembly of screen-printed carbon electrodes, absorption-water pad, nitrocellulose membrane modified by anti-AFP antibody and glass fiber membrane conjugated with ferrocene monocarboxylic acid (FC) labeling AFP. The analytical system utilizes flow-through immunofiltration and competitive immunoassay techniques in combination with an amperometric sensor. The parameters affecting the immunoassay such as selection of filter membrane, membrane pore-size, and antibody binding capacity were investigated and optimized. The immunofiltration system allows us to specifically and directly detect AFP in serum with a low detection limit of 6 ng/mL. The working range is from 6 to 500 ng/mL with an overall analysis time of 5 min for one sample. This electrochemical immunoassay system enabled us to construct a novel point-of-care testing device for the monitoring of biomarker including AFP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.