Abstract

A dual interferometric displacement measurement system is presented where a Wollaston prism interferometer is employed in conjunction with a normal Michelson interferometer. The system operates without the use of external polarizers, apart from those associated with the Wollaston prism interferometer itself. It is shown that an optical path difference induced in the Michelson interferometer can be detected using the Wollaston prism in a normal interferometer arrangement. Further, the interference pattern produced by the Wollaston prism interferometer changes in a measurable, linear fashion as the optical path difference from the Michelson interferometer alters. A simple theoretical analysis of the system is presented and used to derive a computer model of the optical arrangement. Results from an experimental implementation of the system, using a Wollaston prism with a beam separation of 0.5 degrees and a superluminescent diode, of wavelength 825 nm, as a light source, are included and compared to the results from the computer model.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.