Abstract
A modification of the Cary Model 14 for radiometry is described which allows recording in microwatts per square mlllimeter per nanometer on the ordinate and linear wavelength on the abscissa through the range 250-2500 nm. A 5.08-cm diam rotating integrating sphere is used to compare standard to unknown sources at a 30 c/s rate. The undispersed radiation from the sphere is directed through the entrance slit of the monochromator. After dispersion by the monochromator the energy is received by a single phototube or PbS detector as alternate pulses of standard and unknown energy. This design results in common optical elements throughout the instrument except for the standard and unknown sources themselves. An electrical calibration system permits an ordinate scale reading irradiance directly in microwatts per square centimeter per nanometer with simple provisions for range change. The instrument provides a dynamic range of 2 x 10(6) to 1. A large open area is provided for mounting unknown sources. The entrance optical system is designed so that remote sources can be viewed conveniently.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.