Abstract

Using sensor chip immobilized anti-gentamicin monoclonal antibodies in an optical biosensor (BIACORE 3000) resulted in the first single antibody based and label free noncompetitive immunoassay for the direct detection of residues of a low molecular weight compound in a food product. Calibration curves for gentamicin (Mw 466 Da) in buffer and milk showed 50% binding at 20 and 35 ng ml−1, respectively, which is well below the maximum residue limit of 100 ng ml−1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.