Abstract

With the outbreak of COVID-19, it has become very important to improve biosafety measures taken by medical staff. Fewer pretreatment steps correspond to lower chances of infection. The authors established a direct injection technique to analyze levetiracetam (LEV) concentrations in human serum and studied its application in therapeutic drug monitoring. Serum samples were prepared by hollow fiber centrifugal ultrafiltration and the filtrate was directly injected into a ultra-high performance liquid chromatography apparatus (Waters UPLC BEH C18 column: 50 × 2.1 mm, 1.7 μm) for analysis. The mobile phase consisted of acetonitrile and water (8:92) at a flow rate of 1.0 mL/min. The column temperature was maintained at 30°C. The detected wavelength was 210 nm. A linear relationship was obtained for LEV from 0.625 to 80 mcg/mL (r2 = 0.999). The limit of detection for the analysis of LEV was 0.125 mcg/mL. The analysis time was shortened to 4 minutes. The recovery rate of LEV based on the current method was 96.6%-100.1%, whereas the absolute recovery rate was 93.2%-96.8%. The relative SD of intraday and interday precision was <7.3%. Stability was achieved at room temperature for 24 hours after 3 freeze-thaw cycles and at -80°C for 21 days. The method was successfully applied to determine LEV concentrations in the serum of 19 patients. The present method is simple, accurate, and sensitive, and can improve biosafety with the direct injection technique. It is suitable for the analysis of LEV concentrations in therapeutic drug monitoring.