Abstract

A dual direct method for the ultratrace determination of thallium in natural waters by differential pulse anodic stripping voltamrnetry (d.p.a.s.v.) is presented. D.p.a.s.v. at the hanging mercury drop electrode and at the mercury film electrode is used in the concentration ranges 0.5–100 μg Tl l -1, and 0.01–10 μg Tl l -1, respectively. Quantification is aided by the technique of standard additions. The response of the method is optimized for typical natural surface water matrices. An intercomparison of thalium determinations performed by the two anodic stripping methods and electrothermal-atomization atomic absorption spectrometry on normal and thallium-spiked surface water samples demonstrates equivalent accuracy within the range where atomic absorption is applicable. The method appears free from serious interferences.