Abstract
Our objective was to directly compare the diagnostic usefulness of lamellar body counting (LBC) and the TDx-FLM II assay (Abbott Laboratories, Abbott Park, IL) for predicting respiratory distress syndrome (RDS). This was a 5-year, retrospective, cohort study. A diagnosis of RDS was given to infants who received surfactant treatment and/or required ventilator support and/or continuous positive airway pressure for more than 24 hours. There were 172 infants without RDS and 12 with RDS included in the study. By using a TDx-FLM II cutoff of 55 mg/g or more for maturity, the sensitivity was 83%, specificity was 65%, predictive value of a mature result was 98%, and predictive value of an immature result was 14%. These results were similar to LBC using a cutoff of 50,000/microL or more with sensitivity of 92%, a specificity of 60%, a predictive value of a mature result of 99%, and a predictive value of an immature result of 14%. The LBC and TDx-FLM II methods have similar clinical usefulness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.