Abstract

This paper describes the design, construction, and calibration of an electric dipole probe and demonstrates its capability by presenting results on the measurement of electric field excited by a ring electrode in the Large Volume Plasma Device (LVPD). It measures the electric field in vacuum and plasma conditions in a frequency range lying between . The results show that it measures electric field 2 mV cm−1 for frequency . The developed dipole probe works on the principle of amplitude modulation. The probe signal is transmitted through a carrier of 418 MHz, a much higher frequency than the available sources of noise present in the surrounding environment. The amplitude modulation concept of signal transmission is used to make the measurement; it is qualitatively better and less corrupted as it is not affected by the errors introduced by ac pickups. The probe is capable of measuring a variety of electric fields, namely (1) space charge field, (2) time varying field, (3) inductive field and (4) a mixed field containing both space charge and inductive fields. This makes it a useful tool for measuring electric fields in laboratory plasma devices.