Abstract

The hardware and software parts of the signal registration module for a dust-shock mass spectrometer are described. The module allows one to record the spectrum of an ion pulse with frequencies from 100 to 400 MHz; it uses two 14-bit analog-to-digital converters connected to a programmable logic integrated circuit. In this case, a temporary alternation of analog-to-digital converters is used. The test results of the dustproof mass spectrometer at a microparticle accelerator are given (noise level 0.6% of the amplitude of the measured signal, sampling frequency 200 MHz).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.