Abstract

This paper describes an attachment for the Cary model 11 spectrophotometer that will print out the absorbance and wavelength at any number of preselected wavelengths, while the instrument is drawing the usual absorbance—wavelength curve Bidirectional revolut'on counters with electrical readout are used to digitize the wavelength and absorbance data; the digitized data are used to operate a parallel entry printer The program, i.e., wavelengths at which data are to be printed, is controlled by the data from the wavelength digitizer The printer will give absorbance data precise to 0.01 absorbance units when the spectrophotometer is scanning at the rate of 2 1/2 millimicrons/sec.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.