Abstract

We have combined a diffusive sampler and ion chromatograph to determine SO2 in ambient air. The sampler is made of a three‐piece cassette holder containing a triethanolamine‐impregnated glass‐fiber filter. We here describe the absorbent used, sample treatment, method of analysis and the assessment of the sampler in the laboratory and on site. We conclude that the samplers are satisfactory for ion‐chromatographic measurement of SO2, but problems due to loss of SO2 must be considered during long‐term sampling and storage.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.