Abstract

The quality of portal films taken at megavoltage energies is frequently poor, despite the use of various techniques such as lead screens to improve the image. When critical structures, such as the spinal cord, are to be blocked out of the treatment field, it is frequently difficult to determine whether the block is correctly placed. To solve this problem, a diagnostic X ray source has been mounted on the side of a 10 MV linear accelerator to provide accurate verification of patient positioning and location of shielding blocks. The coincidence between the mechanical, 10 MV and diagnostic X ray isocenters is about 1 mm. The system has been designed so that procedures to monitor this coincidence, and adjustment procedures to maintain it, are easily performed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.