Abstract

A patient presented to hospital with chest pain and shortness of breath on 2 occasions 4 weeks apart. Clinical examination revealed an elevated jugular venous pressure consistent with heart failure or elevated filling pressures. The patient was investigated through various modalities including electrocardiogram (ECG), transthoracic echocardiogram, coronary angiography, MRI, cardiac catheterization, positron emission tomography, and an extensive laboratory workup. Serial hs TnI measurements consistently revealed grossly elevated troponin I (>10 000 ng/L). In-lab investigation of increased high sensitivity troponin I (hsTnI) showed evidence of falsely increased troponin due to the presence of heterophilic antibodies. This case demonstrates a complex patient presentation and the value of involving the laboratory medicine team when dealing with potentially discrepant results. This is a rare report of grossly elevated troponin due to heterophilic antibodies for high-sensitivity troponin Abbott assay.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.