Abstract

A practical device ('beam cleaner') is described that selects a specific harmonic (e.g. third order) from other nearby harmonics (e.g. fourth or fifth orders) that are transmitted by the beamline monochromator. The ability to isolate and use a clean harmonic beam effectively extends the energy range of existing beamlines. This device is a tunable medium-resolution X-ray bandpass filter, which consists of a double bent Laue crystal post-monochromator. It is a simple and inexpensive device that does not require precision goniometry and can be operated in air. The device can also be used to passively compensate for vertical beam motion as the primary beamline monochromator is scanned. Experimental results are presented that demonstrate the improvement in XAFS data quality under high harmonic conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.