Abstract
Blood pressure is determined indirectly using a conventional cuff and electric pump for inflation. Cuff pressure is recorded by pressure transducers and oscillograph. The pulse-sensing mechanism is a fluid-filled balloon overlying the brachial artery. This is connected to a pressure transducer and a sensitive oscillographic galvanometer. Systolic and diastolic configurations are easily recognized and their respective pressures determined by referencing the calibrated curve of pressure within the cuff which is recorded simultaneously. This device, with modification, has been successfully used on the human centrifuge and high altitude chambers. Comparison of 90 automatic determinations on 10 subjects with simultaneous auscultatory measures produced a correlation coefficient of 0.97, with the former method showing the greater constancy. This validates the use of this indirect method for selected experimental situations. Submitted on August 5, 1955
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.