Abstract

Scatter is an important problem in computed tomography especially with the increase of X-ray illumination coverage in one single view. Poor scatter correction results in CT HU number inaccuracy, degrades low contrast detectability, and introduces artifacts. Hardware method can be used to handle scatter problem. However, hardware design optimization and scatter correction improvement require an efficient scatter simulation tool. Although Monte Carlo (MC) method can perform precise scatter simulation, simulated noise due to its statistical nature affects the simulation results. In this paper, a deterministic scatter simulation method with radiative transfer equation (RTE) is proposed. Compared to MC method, the deterministic RTE method is free from statistical noise. In order to solve the RTE, a novel iterative spherical harmonics integral formula is developed. Compared to MC method, the results show the accuracy of the proposed method.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call