Abstract
A spectrographic method is presented for the quantitative determination of hafnium in zirconium in the range 0.2%, to 5% HfO2 The sample in the form of a fine powder is burnt in a D.C. arc at 15 amp., 220 V, and the spectrum photographed with a large quartz spectrograph. he hafnia concentration is determined from the ratio of the intensities of the spectral lines HfII 2861.696 Å and ZrII 2856.065 Å. A standard deviation of ± 6% has been obtained by this method for samples essentially free from other impurities, especially Fe, Al, Mg and Na.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.