Abstract

A miniature (200 μm in diameter) cylindrical-core fiber-optic oxygen sensor has been developed for measuring rapid change in oxygen partial pressure (pO <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> ). The fiber-optic sensing element is based on a cylindrical-core waveguide structure formed by coating a thin medical grade polymer sensing film that contains immobilized Pt(II) complexes on silica optical fiber. The performance such as sensitivity and time response of the fiber-optic oxygen sensors were evaluated using luminescence intensity measurement. To determine accurately the response time of the fiber optic oxygen sensors, a test chamber was used to provide rapid changes in the partial pressure of oxygen. The result showed that the time response (time-constant, τ) of this cylindrical-core fiber optic oxygen sensor is less than 50 ms. To our knowledge, this is the fastest such sensor of this size covering the full dynamic range of pO <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> from 0 to 100 kPa.