Abstract

A 0.75 l gridded cylindrical ionization chamber gamma radiation detector using highly purified xenon near the critical point as the detection medium is described. The detector operates at room temperature with a noise subtracted intrinsic energy resolution of 1.8% at 662 keV. The detector design and performance variables are discussed in comparison to previous planar and cylindrical xenon detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.