Abstract

Thermoluminescence (TL) and Radioluminescence (RL) are widely used in dosimetry applications. We present a custom-built integrated system, designated LUMI22, for measuring TL, TL spectroscopy, RL, and RL as a function of temperature. LUMI22 includes a heating system based on Kanthal® A1 alloy (FeCrAl), a microcontroller to regulate the temperature ramps (e.g. 1–5 °C/s). To irradiate samples an X-ray tube (Moxtek 50 kV, 50 μA) is powered, controlled, and monitored by an FTC-200 standard controller. The dose rate at the sample position is 0.43 Gy/min. Light collection includes a Photomultiplier Tube (PMT, Hamamatsu H10493-012:HA, 185–850 nm). Additionally, a miniature fiber optic spectrometer (Ocean Optics, QE65000, range 200–1100 nm) coupled with a 1000 μm diameter fiber optic (QP1000- 2-UV-VIS) was employed for TL and RL spectroscopy measurements. To assess the functionality of the system, it was used to measure TL and RL from Al2O3:C,Mg, Al2O3:C and TLD-100 phosphors which have been previously well investigated. The measured TL and RL data were well compared to the published ones, confirming the functionality of the system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.