Abstract

Methods for separating free and antibody-bound hormone in radioimmunoassays for total triiodothyronine (T-3) and thyroxine (T-4) in unextracted human serum are evaluated. For T-3 assay, a simplified second antibody technique has significant advantages over other methods and gives a mean interassay coefficient of variation of 7.2% over a wide range of values. For T-4 assay, polyethylene glycol is the method of choice and has a mean interassay coefficent of variation of 4.7%. By adding the separating agents initially, the assays are readily semi-automated and may be completed within a working day.