Abstract

Liquid-core waveguides (LCWs), devices that constrain the emitted radiation minimizing losses during the transport, are an alternative to maximize the amount of detected radiation in luminescence. In this work, the performance of a LCW flow-cell was critically evaluated for chemiluminescence measurements, by using as model the oxidation of luminol by hydrogen peroxide or hypochlorite. An analytical procedure for hypochlorite determination was also developed, with linear response in the range 0.2-3.8 mg/L (2.7-51 micromol/L), a detection limit estimated as 8 microg/L (0.64 micromol/L) at the 99.7% confidence level and luminol consumption of 50 microg/determination. The coefficients of variation were 3.3% and 1.6% for 0.4 and 1.9 mg/L ClO(-), respectively, with a sampling rate of 164 determinations/h. The procedure was applied to the analysis of Dakin's solution samples, yielding results in agreement with those obtained by iodometric titration at the 95% confidence level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.