Abstract

Self-expanding metallic stent insertion has been widely applied for the palliative treatment of malignant esophageal strictures. Although it is known as an easy, safe, and effective procedure, complications are well known and include things such as stent migration and esophageal stent occlusion caused by tumor ingrowth. However, metallic stent fractures have been rarely reported in the esophagus, especially for nitinol stents. We report a case of a stent fracture associated with migration in a patient with a malignant esophageal stricture near the gastroesophageal junction. It is highly probable that the stent fracture was due to chemical erosion of the stent caused by gastric juice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.