Abstract
In this paper, a long-term in vitro dynamic calcification of three porcine aortic heart valves was investigated using a combined approach that involved accelerated wear testing of the valves in the rapid calcification solution, hydrodynamic assessment of the progressive change of effective orifice area (EOA) along with the transaortic pressure gradient, and quantitative visualization of the flow. The motivation for this study was developing a standardized, economical, and feasible in vitro testing methodology for bioprosthetic heart valve calcification, which would address both the hydrodynamic performance of the valves as well as the subsequent changes in the flow field. The results revealed the failure of the test valves at 40 million cycles mark, associated with the critical decrease in the EOA, followed by the increase in the maximum value of viscous shear stress of up to 52%, compared to the values measured at the beginning of the study. The decrease in the EOA was subsequently followed by the rapid increase in the maximum streamwise velocity of the central orifice jet up to the value of about 2.8m/s, compared to the initial value of 2m/s, and to the value of 2.2m/s in the case of a control valve along with progressive narrowing of the velocity profile for two test valves. The significance of the current work is in demonstrating a correlation between calcification of the aortic valve and spatial as well as the temporal development of abnormal flow features.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.