Abstract

In this study, a correction method based on background subtraction to remove thermal radiation from luminescent signal was proposed for intensity-based thermometry using thermographic phosphors. Mg4FGeO6:Mn was used as sensor material for a temperature range from 490 to 590 °C. A UV LED was used as excitation source and a spectrometer was used to collect the spectrum of both luminescence and thermal radiation for analysis. Then a CCD camera was used to conduct phosphor thermometry, including calibration and a jet impingement experiment. The calibration results showed that the measurement without correction would fail above 540 °C due to thermal radiation, whereas the correction method could effectively remove the effect of thermal radiation and extend the temperature range up to at least 590 °C. The use of band-pass filter around emission wavelength could reduce the effect of thermal radiation, but correction is still required for the case of higher temperature or lower excitation light. This correction method was then demonstrated in both steady and transient temperature measurements of a jet impingement experiment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.