Abstract
On the basis of analyzed factors that lead to the amplitude variation of an echo signal from identical reflectors in a quality-control sample and a test sample with curved surfaces, the correction factor for the amplitude adjustment is represented as the product of several factors. Calculation methods are proposed for the components of the correction factor for the case of a curved entry surface of an arbitrary shape and a position relative to the transducer.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.