Abstract

I developed and tested a cooled needle electrode (CNE) for radiofrequency (RF) tissue ablation in vitro. A 2-mm needle electrode with two lumina for internal water perfusion and irrigation of the needle tip and a conventional needle electrode were tested in ex vivo calf liver during different levels of output power (wattage). RF lesions produced by the two needle types were further evaluated with a thermal camera. When the CNE was used, a significant increase in the duration of ablation was observed, which caused a significant increase in delivered energy and lesion size when compared with the conventional needle electrode. The largest lesion produced with the CNE was almost spherical and measured 41 x 37 mm (longitudinal x transverse). The cooling effect was reflected in the image obtained with the thermal camera. This technique produced very large lesions compared with conventional methods and may have a role in many different kinds of RF needle ablation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.