Abstract

The present study tested the reliability of an optical scanning device for the objective assessment of postoperative facial swelling. Twenty control subjects bearing a defined volume of water (10-30 mL) in an intraorally carried balloon were tested to assess the measurement accuracy of the device. As a proof of concept, facial volumes of 59 surgical cases were recorded before osteotomy and 1 and 7 days after intervention with the use of a structured light scanner. The median difference between the applied and the measured volumes was 0.67 mL for the control test with the artificial swelling simulated using water balloons. For subjects having third molar osteotomy, extraoral volume increased to 5.29 cm(3) 1 day after surgery (95% CI 5.22-8.52) and decreased to 0.00 mL (95% CI 0.85-2.55) after 7 days. Contact-free visible-light 3-dimensional scanning is reliable for the objective assessment of postoperative facial swelling.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.