Abstract

A filtered backprojection reconstruction algorithm was developed for cardiac single photon emission computed tomography with cone-beam geometry. The algorithm reconstructs cone-beam projections collected from ;short scan' acquisitions of a detector traversing a noncircular planar orbit. Since the algorithm does not correct for photon attenuation, it is designed to reconstruct data collected over an angular range of slightly more than 180 degrees with the assumption that the range of angles is oriented so as not to acquire the highly attenuated posterior projections of emissions from cardiac radiopharmaceuticals. This sampling scheme is performed to minimize the attenuation artifacts that result from reconstructing posterior projections. From computer simulations, it is found that reconstruction of attenuated projections has a greater effect upon quantitation and image quality than any potential cone-beam reconstruction artifacts resulting from insufficient sampling of cone-beam projections. With nonattenuated projection data, cone beam reconstruction errors in the heart are shown to be small (errors of at most 2%).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.