Abstract

An automatic magnetic resonance spectrometer with a pulsed magnetic field is described. It has the following characteristics: a field range of 0–100 kOe, a frequency band of 25–140 GHz, a temperature range of 4.2–300 K, and a pulse duration of 12.6 ms. The spectrometer is automated to the CAMAC standard using standard modules. To record a magnetic-resonance spectrum, the following procedure is applied: the resonance-absorption signal is determined by a 10-bit analog-to-digital converter, and the magnetic filed is scanned by simulating a current pulse in a discharge circuit of the solenoid using a single input parameter—the initial charge voltage of the capacitor bank. In calibration measurements performed with spherical yttrium–iron-garnet samples, the average measurement error was 0.15%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.