Abstract

Accurately positioning components during radial head arthroplasty is difficult due to the complex shape of the radial head. The objective of this study was to develop a navigation system for patient-specific radial head implants using intra-operative surface-based registration techniques. Using an optical tracking system, an implant stem was navigated using real-time visualisation of the implant position and orientation (n = 8). Absolute positioning errors in the ML and AP directions were 0.8 ± 0.5 mm (and 1.1 ± 0.9°) and 0.8 ± 0.6 mm (and 1.2 ± 1.7°), respectively. Registration error was less than 1.0 mm. The current system is an important first step in the development of a computer-assisted system for implant navigation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.