Abstract
Measurements show the detection efficiency of the NaI(Tl) detector depends on the source-detector configuration. In this work, the geometrical and intrinsic detection efficiencies of the 2in both in height and diameter and 3in both in height and diameter cylindrical NaI(Tl) detectors have been calculated for low-energy gamma ray. In these calculations, the effects of cover layers of NaI(Tl) (aluminum housing and reflector) have been taken into consideration. To evaluate the calculation model, the detection efficiency of the detector is measured and modelled using MCNP simulation tool. Isotropic 241Am was used as a point source. Results show the calculation model is in good agreement with the experiment and Monte Carlo simulation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.