Abstract

A comprehensive orthogonal two-dimensional liquid chromatography (2D-LC) based on the modification of mobile phases was developed with a sample loop–valve interface. To improve the compatibility of mobile phases and analysis speed, some special solvents were chosen as the mobile phases, and the column temperature was elevated to decrease the viscosity of mobile phases of reversed-phase liquid chromatography (RPLC). Based on this principle, the first dimension was normal-phase liquid chromatography (NPLC) with a SiO 2 column, and the second dimension was reversed-phase liquid chromatography containing two tandem C18 columns. 1,4-Dioxane was used in the NPLC mobile phase, and isopropyl alcohol was employed in the RPLC mobile phase. Moreover, the elevated column temperature enabled the reduction of the backpressure and using tandem C18 columns to improve the resolving power in RPLC. The new comprehensive 2D-LC system and applied strategy offered a novel idea for construction of 2D-LC system. A traditional Chinese medicine, Zhengtian pill, was used as the test sample to evaluate the constructed 2D-LC system. 876 peaks were detected, and the peak capacity reached 1740.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.