Abstract

The "prescan" process for setting the gain of an MRI scanner can be confounded by an effect arising from the spatial variation of the angle between the vector reception field of a local or surface coil and the excitation field of a whole-body coil. It is most pronounced when the phase-encoding direction is parallel to the plane of the coil but not along the magnetic field. The largest signal is shown to reside in a data row that is shifted from the one that corresponds to zero phase-encoding gradient. The effect depends on the field-of-view and the dimensions of the receiving coil. Strategies are suggested by which manufacturers of scanners can cope with this intrinsic problem. In the meantime, users of surface coils should be aware of the difficulty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.