Abstract

This work describes an immunosensor for the determination of okadaic acid (OA) using a Love-wave sensor with the immunogold staining method. In this assay, OA-BSA was captured by the anti-BSA antibody (BSA-MAb) bond with staphylococcal protein A (SPA) on the sensor surface via Fc regions. Cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were employed to monitor the fabrication steps of immunosensor using the redox probe of K3Fe(CN)6. An indirect competitive immunoassay format was employed to detect OA after the immunosensor was prepared and gold nanoparticles (AuNPs) were introduced to amplify the phase signal. The results demonstrated that developed immunosensor was able to specifically respond to OA ranging from 10 to 150ng/mL, providing limit of detection of 5.5ng/mL. When applied to determine OA concentration in real sample, the results agreed well with those obtained by the commercial ELISA assay. The developed immunosensor is therefore suitable and has great application potential in detection of OA in shellfish.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.