Abstract
Several recently developed tests of visual function, including the Potential Acuity Meter (PAM), laser interferometer (LI), white-light interferometer (WLI), blue field entoptic phenomenon, and focal electroretinogram (ERG) were compared in 81 eyes with clear media and known macular disease. The results indicate that the PAM, the LI, and the WLI overreact relative to Snellen acuity. Laser interferometric acuity values differed from Snellen acuity by at least 1.5 octaves in approximately 40% of all eyes, regardless of stimulus size (2, 5, or 8°). Similar results were obtained with the WLI. Agreement with Snellen acuity was better for the PAM, with 91 % of eyes falling within 1.5 octaves of Snellen acuity. Blue field and focal ERG results were categorized as normal or abnormal. While not producing Snellen equivalents, abnormal results from the blue field and focal ERG corresponded with poor Snellen acuity (<20/40) in 65% and 91 % of eyes, respectively. Assuming that media opacities do not prevent adequate retinal stimulation, the present results suggest that the PAM and focal ERG are the most reliable for evaluating macular function when maculopathy is present.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.