Abstract

A sample of known elemental concentrations was activated in the bremsstrahlung photon beam which was created by a pulsed electron LINAC. Several procedures of photon activation analysis, including those applied with/without reference material and with/without photon flux monitor, were conducted to make a comparison of their precision and accuracy in practice. Experimental results have indicated that: (1) relative procedures usually produce better outcome despite that the absolute measurement is straightforward and eliminate the assistance of reference materials; (2) among relative procedures, the method with internal flux monitor yields higher quality of the analytical results. In the article, the pros and cons of each procedure are discussed as well.