Abstract

The accurate measurement of low serum creatinine levels is necessary for estimating clinically useful creatinine clearances in the pediatric population. This study compares two routine clinical methods: the kinetic Jaffé with the newer Kodak enzymatic method against our reference method, high-pressure liquid chromatography, for the measurement of serum and urine creatinine levels in neonates. One hundred and twenty-five serum and 59 urine creatinines and 56 absolute creatinine clearances were measured in neonates ranging from 23 to 46 weeks (mean 32 weeks) post-conceptional age and weighing 480-4398 g (mean 1650 g). Urine creatinine levels, and serum creatinine levels greater than 0.8 mg/dl were equivalent for both clinical methods. However, the enzymatic method was much more accurate (P less than 0.001) than the kinetic Jaffé method for serum creatinine measurements of less than or equal to 0.8 mg/dl. We conclude that the enzymatic methodology is a better clinical choice for the accurate measurement of serum creatinine levels when using these values for the determination of neonatal renal function.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.