Abstract

Measurement setups that characterize the impedance of suspensions of blood over the wide frequency range of 100 Hz to 100 MHz are presented in this paper. The performance of the two- and four-electrode techniques have been compared and evaluated. By applying a combination of the two measurement techniques the best result is achieved when taking into account the main nonidealities, such as electrode polarization impedance and parasitic capacitances. It has been found that the conventional three-element model for the impedance of blood can be used for frequencies up to 1 MHz. For frequencies exceeding 1 MHz, an extended model is introduced where a constant phase angle element is used for modeling the cell membrane and a capacitor C(liq) is added for modeling the electrical capacitance of water in blood.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.